2.2 Installation Environment and Mounting Space

2.2.1 Installation Environment

Avoid Installing the Equipment in the Following Types of Locations:
• Locations with environmental temperature that exceed a range of 0 to 50 °C.
• Locations with a relative humidity exceeding a range of 35%RH to 75%RH.
• Locations with a condensing due to a sharp temperature change.
• Locations with an atmosphere containing corrosive gases or flammable gases.
• Locations with a lot of fine particles, iron filings or salt.
• Locations with an atmosphere likely to contain organic solvents (such as benzene, paint thinner), and
alcohol as well as strongly alkaline materials (such as ammonia and caustic soda).
• Locations subject to direct vibration or impact.
• Locations subject to direct sunlight.
• Locations where the device can become wet with water, oil or chemicals.
• Locations where the device is applied with excessive load.
Noise Countermeasure
• Keep the AX40 away from the devices that generate large switching surges, such as high-voltage
wires, high-voltage devices, power-driven lines and power-driven machinery as much as possible.
• Keep the AX40 away from the devices with the transmission system such as radio equipment as
much as possible.
Countermeasure for Heat Radiation
• Do not mount the device on the object with strong heat sources such as the heater, transformer and
high-capacity resistor.
• When installing, set the device in the following direction for heat radiation.