Configure the Transmitter battery status report for viewing and printing
2001989-203C
To disable the Remote Service Agent, select Disable.
5.
Select Save to save your changes.
Verify Remote Service startup
Check for any system log entries indicating a remote service startup failure.
ApexPro log files on page 6-22.
Test connectivity to GE back office
Authorized GE service personnel must contact the appropriate service organization to
establish connectivity. Testing connectivity is the responsibility of the appropriate
service back office organization in your region.
NOTE
This report is available if the system language is set to English.
CIC Pro v4.0.7 or greater is required.
The Transmitters battery status report is a summary of battery life for the ApexPro
FH transceivers as well as ApexPro CH, ApexPro, and NEC transmitters. This report
is available via Webmin and is printable for all admitted, supported telemetry
transmitters.
NOTE
For this report to function, there must be at least one ATS existing on the
network.
The report displays the following information:
Care unit and bed name
TTX ID and battery level, displayed as a percentage of the full charge remaining
NOTE
To provide the clinician ample time to react, the battery level will
change to Change Battery at a battery level no lower than 10%. The
Change Battery text is printed in bold red letters so that it stands out on
the report.
When a low battery condition occurs, approximately 1-2 hours before
the telemetry battery dies, a LOW BATTERY message is displayed at
the CIC Pro center with no audio alarm. When the battery dies, an
audible system warning alarm will be generated at the CIC Pro center.
Webmin provides the ability to sort the information by any column on the report. The
report can be viewed and printed by creating a browser favorite at the CIC Pro center
that points to an ATS tower.
FE creates a favorite at the CIC Pro center
NOTE
Consult with the hospital staff to determine which CIC Pro center needs to be
configured with this feature.
ApexPro™ Telemetry
Configuration
See View
5-13