Electromagnetic Compatibility (Emc; Guidance And Manufacturer's Declaration; Electromagnetic Emissions - GE ApexPro Service Manual

Telemetry server
Hide thumbs Also See for ApexPro:
Table of Contents

Advertisement

Electromagnetic compatibility

Electromagnetic compatibility (EMC)

Guidance and manufacturer's declaration

Electromagnetic emissions

Emissions test
RF Emissions
EN 55022
RF Emissions
EN 55022
Harmonic Emissions
EN 61000-3-2
Voltage Fluctuations/
Flicker Emissions
EN 61000-3-3
D-2
Changes or modifications to this system not expressly approved by GE could cause
EMC issues with this or other equipment. This system is designed and tested to
comply with applicable regulation regarding EMC and needs to be installed and put
into service according to the EMC information stated in this appendix.
CAUTION
Use of portable phones or other radio frequency (RF) emitting
equipment near the system may cause unexpected or adverse
operation.
CAUTION
The equipment or system should not be used adjacent to, or stacked
with, other equipment. If adjacent or stacked use is necessary, the
equipment or system should be tested to verify normal operation in
the configuration in which it is being used.
The ATS is intended for use in the electromagnetic environment specified below. It is
the responsibility of the customer or user to assure that the ATS is used in such an
environment.
Compliance
Group 1
The device uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
Class A
Class D
Mains power quality should be that of a typical commercial and/or hospital
Complies
environment.
ApexPro™ Telemetry
Electromagnetic environment – guidance
2001989-203C