GE ApexPro Service Manual page 127

Telemetry server
Hide thumbs Also See for ApexPro:
Table of Contents

Advertisement

PN 2021518-001
PN 2021518-001
2001989-203C
NOTE
Before replacing the RAID card, make sure the ATS is running the proper
version of the ATS OS image.
NOTE
If the software version currently running is earlier than the software included
in the reload kit, this operation may result in a software upgrade.
5.
Complete the preparation procedures.
7-3.
6.
Remove the chassis cover.
7.
Disconnect the hard drive ATA serial cables from the RAID card.
8.
If you are replacing the RAID card, remove the ATA serial cables from the
connection ports at the back of the hard drives, and discard the cables. The RAID
card FRU kit contains replacement ATA serial cables as well.
9.
Disconnect the LED connector from the front edge of the RAID card.
10. Remove the 3 screws securing the back of the CPU module to the chassis.
11. Remove the 2 screws securing the front of the CPU module to the chassis.
ApexPro™ Telemetry
Parts lists, drawings and replacement
See Field replacement preparation on page
See Chassis cover on page 7-6.
PN 2021518-002/2021518-003
PN 2021518-002/2021518-003
014A & 251A
049A & 252A
046A
7-9

Advertisement

Table of Contents
loading

Table of Contents