Digital Blood Pressure Monitor
Model UA-1020
Instruction Manual
Original
'
Manuel d
instructions
Manual de Instrucciones
Manuale di Istruzioni
使用手冊
1WMPD4002224A

Advertisement

Table of Contents
loading

Summary of Contents for AND UA-1020

  • Page 1 P.M.S(Instruments) Ltd, Waldeck House, Waldeck Road, Maidenhead, SL6 8BR Tel 01628 773233 www.pmsinstruments.co.uk Digital Blood Pressure Monitor Model UA-1020 Instruction Manual Original ’ Manuel d instructions Manual de Instrucciones Manuale di Istruzioni 使用手冊 1WMPD4002224A...
  • Page 2: Table Of Contents

    WHO Classification Indicator ..........15 About Blood Pressure............16 What is Blood Pressure? ..........16 What is Hypertension and How is it Controlled?....16 Why Measure Blood Pressure at Home?......16 WHO Blood Pressure Classification ......... 16 Blood Pressure Variations..........16 Troubleshooting ..............
  • Page 3: Dear Customers

    Extremes in temperature, humidity, direct sunlight, shock or dust should be avoided. Clean the device and cuff with a dry, soft cloth or a cloth dampened with water and a neutral detergent. Never use alcohol, benzene, thinner or other harsh chemicals to clean the device or cuff.
  • Page 4: Parts Identification

    MEMORY I.H.B. Symbol (Irregular heartbeat symbol) Cuff Fit Error Symbol Systolic Pressure Movement Error Symbol TriCheck Symbol Diastolic Pressure Heart Mark WHO Classification Indicator and Pressure Bar Indicator Pulse Rate Battery Indicator Clock Display AM / PM Mark English 3...
  • Page 5: Symbols

    Standby and Turn the device on. Battery installation guide Direct current Serial number Date of manufacture 2010 Type BF: Device, cuff and tubing are designed to provide special protection against electrical shocks. EC directive medical device label WEEE label Manufacturer...
  • Page 6 Take another measurement. Unstable blood pressure due to Remain still during movement during measurement. measurement. The systolic and diastolic values are within 10 mmHg of each other. The pressure value did not increase Apply the cuff correctly, during the inflation.
  • Page 7: Operation Mode

    Operation Mode 1. Normal Measurement Press the START button. Blood pressure is measured and the data is stored in memory. This device can store the last 90 measurements in memory. 2. Recalling the Data Press the button to recall the data in memory.
  • Page 8: Using The Monitor

    When (LOW BATTERY mark) blinks on the display, replace all batteries with new ones. Do not mix old and new batteries. It may shorten the battery life, or cause the device to malfunction. Replace the batteries two seconds or more after the device turns off.
  • Page 9: Adjusting The Built-In Clock

    2. Select the year using the button. Press the SET button to set the current year and move to month/day selection. The date can be set anywhere between the years 2010 Month and 2059. 3. Select the month using the button.
  • Page 10: Applying The Arm Cuff

    Relax for about five to ten minutes before measurement. Place the center of the cuff at the same level as your heart. Remain still and keep quiet during measurement. Do not measure right after physical exercise or a bath. Rest for twenty or thirty minutes before taking the measurement.
  • Page 11: Measurements

    Pressure” on the next page. When the measurement is complete, the systolic and diastolic pressure readings Measurement and pulse rate are displayed. in progress The cuff exhausts the remaining air and deflates completely. Systolic pressure Press the button to turn the device off. START...
  • Page 12: Measurement With The Set Pressure

    Measurements Model UA-1020 is designed to detect the pulse and to inflate the cuff to a systolic pressure level automatically. If re-inflation occurs repeatedly, use the following methods. Measurement with the SET Pressure During the blood pressure measurement, re-inflation may occur.
  • Page 13: Notes For Accurate Measurement

    An individual's blood pressure varies constantly, depending on what you are doing and what you have eaten. What you drink can have a very strong and rapid effect on your blood pressure.
  • Page 14 After one minute, the second First measurement measurement starts. When the measurement is complete, the average readings of the first and second measurements are displayed, then a one-minute countdown starts for Results of the the third measurement. first measurement...
  • Page 15: Recalling The Memory Data

    Recalling the Memory Data Note: This device stores the last 90 measurements in memory. 1. Press the button. Press The average of all measurements and the number of data are displayed. Average systolic (If no data, “0” is displayed. Press the Average diastolic...
  • Page 16: What Is An Irregular Heartbeat

    25% from the average of all heartbeats during the blood pressure measurement. It is important that you are relaxed, remain still and do not talk during measurements. Note: We recommend contacting your physician if you see this symbol frequently.
  • Page 17: About Blood Pressure

    30 to 50 mmHg due to various conditions during the day. In hypertensive individuals, variations are even more pronounced. Normally, the blood pressure rises while at work or play and falls to its lowest levels during sleep. So, do not be overly concerned by the results of one measurement.
  • Page 18: Troubleshooting

    The cuff position is not Place your arm on a table with your too high or too correct. palm facing upward and the cuff at low. the same level as your heart. If you have a very weak or irregular heat beat, the device may have difficulty in determining your blood pressure.
  • Page 19: Maintenance

    Maintenance Do not open the device. It uses delicate electrical components and an intricate air unit that could be damaged. If you cannot fix the problem using the troubleshooting instructions, contact the authorized dealer in your area or our customer service department. The A&D customer service will provide technical information, spare parts and units to authorized dealers.
  • Page 20 A&D Company, Limited 1-243 Asahi, Kitamoto-shi, Saitama 364-8585 JAPAN Telephone: [81] (48) 593-1111 Fax: [81] (48) 593-1119 A&D INSTRUMENTS LTD. Unit 24/26 Blacklands Way, Abingdon Business Park, Abingdon, Oxfordshire OX14 1DY United Kingdom Telephone: [44] (1235) 550420 Fax: [44] (1235) 550485 A&D Australasia Pty Ltd.