Performing a UDDS
calibration
Alterations
3-18
UDDS calibration is used to calibrate the ultrasonic document detection
system for multifeeds. UDDS calibration is rarely needed. This calibration
should only be performed if there is a feeding or multifeed detection issue.
1. Select UDDS Calibration. The UDDS Calibration screen will be displayed.
2. You will be prompted to place an A4 size 75-80 g/m
bond sheet of paper in the input elevator in portrait orientation.
3. Press the Start/Pause button. The calibration results will be displayed.
This option is for Service Use only. Only use this option when directed to do so
by Kodak Alaris Service personnel.
2
or lettersize 20 lb.
A-61837 June 2015