E. How to disable battery when system hang
up –CyberMed H22/N22
1. When system hang up ,press power button 4 seconds to turn
off system . If you can't power on system by power button.
Please follow below step to reset system.
2. Remove AC power cord, then the battery LED will on.
3. Press volume up key 10 seconds, then all LED will turn on and
turn off
4. Plug in AC power cord again, press power button then you can
power on system.
54