Stage Rotation

3.5
Stage Rotation
3.5.1
To rotate the circular graduated stage, loosen the
stage rotation clamp screw and turn the whole
stage carefully by hand.
The angle of rotation can be read to 0.1 degrees
with the two vernier scales.

Attachable Mechanical Stage (Option)

3.5.2
The optional attachable mechanical stage is
installed by inserting the two pins on the bottom
into the two pinholes on the stage surface. Tighten
the clamp screw using the supplied hexagonal
wrench.
To move the specimen position under observation,
rotate the knob on the stage. You can adjust the
position in the X-direction and Y-direction
individually. (Travel range: 35 x 25 mm) The travel
amount can be read to 0.1 mm with the two
vernier scales.
Specimens are fixed with a lever. So you can
change specimens with easy operation.
Chapter 3 Individual Operations
3.5 Stage Rotation
36
Vernier
Stage rotation
clamp screw
Clamp screw
Positioning pins