Pump Will Not Prime; Leakage At Tubing; Failure To Run; 3. Troubleshooting - Milton Roy LMI P1 Basic Manual

Electronic dosing pumps
Table of Contents

Advertisement

IV – 3. TROUBLESHOOTING

Pump will not prime

Pump not turned on or plugged in.
Turn on pump/plug in pump.
Output dials not set properly.
Always prime pump with speed at 80 % and stroke
at 100 %.
Foot valve not in vertical position on bottom of tank.
See Chapter II – 1. Section Foot valve/ceramic
weight.
Pump suction lift too high.
Maximum suction lift is 1,5 m. Pump with High
viscosity liquid handling assemblies require flooded
suction.
Suction tubing is curved or coiled in tank.
Suction tubing must be vertical. Use ceramic weight
supplied with pump. (see Chapter II – 1. Section
Foot valve/ceramic weight).
Fittings are over tightened.
Do not overtighten fittings. This causes seal rings to
distort and not seat propertly which causes pump to
leak back or lose prime.
Air trap in suction valve tubing.
Suction tubing should be as vertical as possible.
AVOID FALSE FLOODED SUCTION (see Chapter
II – 1. Section Flooded suction).
Too much pressure at discharge (pump without 4-
function valve).
Shut off valves in pressurized line. Disconnect
tubing at injection check valve (see Chapter III – 1.
Priming). When pump is primed, reconnect
discharge tubing.
Pump loses prime
Solution container ran dry.
Refill container with solution and reprime (see Part
III).
Foot valve not in vertical position on bottom of tank.
See Chapter II – 1. Section Foot valve/ceramic
weight.
Pump suction lift too high.
Maximum suction lift is 1,5 m. Pump with High
viscosity liquid handling assemblies require flooded
suction.
Suction tubing is curved or coiled in tank.
Suction tubing must be vertical. Use ceramic weight
supplied with pump. (see Chapter II – 1. Section
Foot valve/ceramic weight).
Fittings are over tightened.
Do not overtighten fittings. This causes seal rings to
distort and not seat propertly which causes pump to
leak back or lose prime.
Air trap in suction valve tubing.
Suction tubing should be as vertical as possible.
AVOID FALSE FLOODED SUCTION (see Chapter
II – 1. Section Flooded suction).
LMI, Doseur à membrane, L001, Rev.B, 1098
Air leak on suction side.
Check for pinholes, cracks. Replace if necessary.

Leakage at tubing

Worn tubing ends.
Cut tubing about 25 mm (1'') off tubing and then
replace as before.
Loose or crack fitting.
Replace fitting if cracked. Carefully hand tighten
fittings. Do not use pipe wrench. Once fitting comes
into contact with seal ring, tighten an additional 1/8
or ¼ turn.
Worn seal rings.
Replace balls and seal rings (see part IV).
Solution attacking Liquid handling assembly
material.
Consult Technical Assistance Department for
alternate materials compatible with the pumped
solution.
Low output or failure to pump against
pressure
Pump's maximum pressure rating is exceeded by
injection pressure.
Injection pressure cannot exceed pump's maximum
pressure. See pump data plate.
Worn seal rings.
Replace them : see Chapter IV –1.
Ruptured diaphragm.
Replace diaphragm : see Chapter IV –1. Section
Diaphragm replacement.
Incorrect stroke length.
Check zero on pump/Re-zero pump (See Chapter
IV – 2.).
Tubing run on discharge may be too long.
Longer tubing runs may create frictional losses
sufficient to reduce pump's pressure rating.
Clogged foot valve strainer.
Remove foot valve strainer when pumping slurries
or when solution particles cause strainer to clog.

Failure to run

Pump not turned on or plugged in.
Turn on or plug in pump. Check also programming
(A9, B9, C9 Series).
EPU failure.
Disassemble pump and measure the resistance of
the EPU across the EPU wires. Resistance reading
should be in accordance to the table (see Chapter
IV – 4.). Also check EPU leads to ground.
Pulser failure.
The pulser should be replaced if EPU checks out
OK.
IV - 3

Advertisement

Table of Contents
loading

Table of Contents