Laboratory glassware (LG)

Wide-necked laboratory glassware, e.g.
beakers, wide-necked Erlenmeyer
flasks and Petri dishes, and cylindrical
shaped items e.g. test tubes, can be
cleaned and rinsed, internally and
externally, by the action of the rotating
spray arms. They can be arranged in
full, half and quarter inserts in an empty
mobile unit with spray arms.
Narrow-necked items e.g. narrow-
necked Erlenmeyer flasks, conical
flasks, measuring flasks and pipettes,
require mobile injector units or injector
modules.
The following instructions relate only to
basic preparation and loading of
glassware.
Loading the machine
– Petri dishes and similar should be
placed in the correct insert with the
soiled side facing the centre.
– Pipettes should be placed with the
tip facing downwards.
– Quarter inserts should be placed as
close to the middle of the unit as
possible.
– Glass items must not get in the way
of the door opening.
Areas of application
Preparing the load
^ Empty all glassware before loading
into the machine (paying particular
attention to regulations regarding
epidemics).
,
Ensure that acids and residual
solvents, especially hydrochloric
acid or chlorides, cannot get into the
wash cabinet.
^ Remove all agar residues from Petri
dishes.
^ Remove blood clots and residues
from test tubes, etc.
^ Remove all stoppers, corks, labels,
sealing wax residues, etc.
^ Small parts - such as stoppers and
taps - should be secured in suitable
basket inserts.
E 940 Mobile injector unit
On two levels, for narrow necked
laboratory glassware (can also be used
without the upper level).
35