Areas of application

This washer-disinfector can be fitted
with a variety of mobile units, which can
be equipped with a variety of modules
and inserts depending on the type and
shape of items requiring cleaning and
disinfection.
Select mobile units, modules and
inserts which are appropriate for the
application.
Notes on the individual areas of
application and examples of loading
are given on the following pages.
Before starting a programme you
should carry out a visual check on
the following:
– Is everything correctly
loaded/connected for cleaning?
– Are the spray arms clean and do
they rotate freely?
– Are the filters clean? Remove any
coarse soiling and clean them
properly if necessary.
– Is the adapter connecting the water
supply to the spray arms/jets
correctly connected?
– Are all chemical containers
sufficiently filled?
28
At the end of each programme:
– Carry out a visual check of the load
for cleanliness.
– Check that all hollow shafted
instruments are still securely located
on their jets.
,
Any hollow instruments that have
become disconnected from their
adapters during reprocessing must
be reprocessed again.
– Check that the lumen of hollow
instruments are free of obstruction.
– Check that jets and connectors are
securely held in position in the
baskets or inserts.

Protein test

Cleaning results should be subjected to
periodic protein tests, e.g. with the
Miele test kit.