Master switch

The master switch disconnects the
operator side of the washer-disinfector
from the electricity supply.
^ Set the master switch to I ON so that
the machine can be operated.
The washer-disinfector is ready to use
at the end of the start-up procedure.
The following message may appear:
"New mains frequency 60Hz Flow
rate automatically adjusted" or "New
mains frequency 50 Hz Flow rate
automatically adjusted.".
During the start-up procedure, the
profitronic controls have detected a
different mains frequency and
adjusted the flow rate of the
dispensing pumps accordingly.
Before the machine is operated, a
check must be made to ascertain if
the change in mains frequency was
caused by an interruption to the
power supply.
In this case, the wrong quantity of
agent would be dispensed during the
programme.
The fault message can only be
confirmed in operating level D or
Service mode.
46

Switching on

^ Press button s for at least 1.5
seconds.
The length of time the s button can
be pressed is set at a maximum of
10 seconds (see Programming
manual "System functions / On/Off
button activation time").
Depending on which operating level
has been set, the following will appear
in the display:
Operating
Display shows
level
A and B
A list of free access
programmes
C
Instructions on how to
operate the automatic
mobile unit recognition
D
The following options:
– Programme overview
– Programming
– Settings J
The display will switch off
automatically after approximately 10
minutes when the washer-disinfector
is not in operation.