Error Message - Fora D15 User Manual

Blood glucose plus blood pressure monitoring system
Table of Contents

Advertisement

Error Message

MESSAGE
Err.00
Err.02
Err.04
Err.07
Err.08
Err.22
Err.24
Err.25
Err.26
Err.05
Err.06
Err.09
Err.10
Err.11
Err.15
Err.20
Err.28
52
CAUSE
Weak pulse.
Monitor can not determine
systolic pressure.
Monitor can not determine
systolic or distolic pressure
or it is out of range.
Invalid coding.
Used strip inserted.
Ambient temperature is
below operating
temperature.
Ambient temperature is
over operating
temperature.
Those messages indicate internal problems with the
monitor. Do not attempt to fix them.
Please contact Customer Care Line at 1-866-469-2632.
WHAT TO DO
Refit cuff tightly, relax,
and repeat measurement
as shown in the section,
"Correct Application of the
Pressure Cuff ".
Try a new code strip and
insert again.
Use a new strip.
Please operate this monitor
between 50°F and 104°F
(10°C and 40°C).