Replacing Legrest Actuators - Invacare STORM Service Manual

Hide thumbs Also See for STORM:
Table of Contents

Advertisement

14.3 Replacement of Actuators for Legrests
Disassembly
• Pull the plug of the actuator to be replaced out of the
socket.
• Remove the cap of the upper actuator fastening.
• Now disassemble the actuator by removing two fastening
screws (SW 13 mm).
Assembly
• Install the new actuator and fasten it by tightening the
two fastening screws.
• Lead the actuator cable to the corresponding socket board and
plug the plug into the socket.
• Conduct a performance test.

NOTE:
• Do not tighten the fastening screws too tightly.-
The actuator must be able to move within
its fastenings.
• Install the cable in a way where it cannot get
pinched.
76
Actuator for Legrest
Fastening Screw
M8 x 35 - SW 13 mm
Fastening Screw
M8 x 35 - SW 13 mm
Actuator,
Upper Actuator Fastening
M8 x 35 - SW 13 mm
Washer
Tapped
bushings
for legrests
Washer
M8 x 35 - SW 13 mm
Lower Actuator Fastening
Fastening screws
Cap
Fastening screw

Hide quick links: