Phase To Neutral (Or Phase To Phase) Testing - Megger MFT1700 series User Manual

Mft1700 series multifunction testers
Hide thumbs Also See for MFT1700 series:
Table of Contents

Advertisement

.4
Press the Function key <-> to select the "2Lo" mode.
.5
Press TEST to start the test sequence.
.6
On completion of the test, the display will show the loop resistance on the large display segments, and the fault
current on the small display segments.
7.3

Phase to Neutral (or Phase to Phase) testing

Note: Only the "2Hi" mode is available on this range.
.1
Set the LEFT rotary range knob to the
.2
On the MFT1720 and 1730, set the RIGHT rotary knob to
MFT1710
.3
Connect the test leads to the circuit as below, with the Red test lead connected to the L1 (Red terminal on the MFT
and the Blue test lead to the Blue (L3) terminal.
 
 
 
 
range.
 
MFT1720 and MFT1730
 
.
 
 
19