How To Install Weighing Mechanism - KERN EW Service Manual

Ew / ew-c / eg (n) series electronic precision balances
Table of Contents

Advertisement

* For EG series (With-internal-calibration type), refer to section 5-6.
(4) Remove four mechanism unit fixing screws
(5) Remove (cut) the board band.
(6) Remove the mechanism unit by holding at the chassis.

5.3.2 How To Install Weighing Mechanism

(1) Place the mechanism on the chassis. Fit it with the fixing screws.
(2) Solder wires of the Tuning-fork sensor to the AJDP circuit board.
(3)
Cover the mechanism unite with the sensor cover. Fit it with four
(1)
fixing screw. (EW 120-1200).
Fix the Tuning-fork sensor wire to hexagon pillar with cord
(2)
band. (EW2200-12K)
(4) Cover the scale referring to 5-2.
Circuit Board Fixing Screw
- 33 -
EW_EW-C_EG_N-SH-e-0722