Download Print this page

Display Error Codes - Rice Lake SURVIVOR CW-80 Operating Manual

Checkweigher
Hide thumbs Also See for SURVIVOR CW-80:

Advertisement

7.

Display Error Codes

Display
X
0
1
2
3
4
5
6
7
14
CW-80 Supervisor's Operating Guide
Condition
OVERLOAD
Weight on scale is greater than calibrated capacity
UNDERLOAD
Displayed weight is less than calibrated capacity range.
ID ILLEGAL
ID number 0 cannot be stored. ID number must be a
positive integer.
NO CALIB.
Configuration settings relating to scale capacity, divi-
sions, grads, or decimal pt. have been changed.
Recalibrate scale using current settings.
No tolerances have been entered for the displayed ID
NO TOLERANCE
BUFFER FILLING
Buffer 90% full.
BUFFER FULL
Buffer 100% full.
FAIL SIGNAL
No load cell signal present.
DATA OUTPUT
Cannot display lb/oz data in 6 display digits.
ERROR
CALIBRATION
CAL SP value must be greater than 20% of full scale.
ERROR
LOW SIGNAL
Less than 0.3 µV/grad signal is present.
ERROR
Keypad button X is not producing a signal.
KEYPAD BUTTON
ERROR
Keypad button 0 = target error.
Keypad button 1 = print error.
Keypad button 2 = zero error.
Keypad button 3 = over error.
Keypad button 4 = ID error.
Keypad button 5 = tare error.
Keypad button 6 = units error.
Keypad button 7 = under error.
EEPROM checksum failure. CW-80 has defaulted to
NO UNITS
original units.
ID ERROR
EEPROM write failure. Unable to write to ID selected.
Description

Advertisement

loading