Introduction This manual is intended for use by qualified service technicians responsible for installing and servicing the CW-90/CW-90X checkweighing scale. Ma nuals can viewe d or downlo aded from the Rice Lake Weighing Syste ms website at www.ricelake.com/manuals Safety Safety Symbol Definitions:...
Overview The CW-90/CW-90X is a high-speed digital weight indicator and scale base programmed to compare weight readings with predetermined tolerance limits defining an “accept” band. If the current weight reading is within the acceptable range, the green “ACCEPT” LED is illuminated. If the current weight reading is less than the acceptable range, one or more of the red “UNDER”...
Allows the display of the current “under tolerance” value, or allows setting the current “under tolerance” value. When the MODE parameter is set to TARGWT or TARG% (see Section on page 19), this key is used to acquire a weight value and assign it as the desired “target” value. The CW-90/CW-90X then computes the final TARGET over/under values using the target value and the over/under tolerance settings defined during the weighing operation.
Indicates that a preset tare weight has been keyed in or entered via the EDP serial port. Displays which unit of measure is being used. % LED Percent mode kg LED Kilograms g LED Grams lb LED Pounds oz LED Ounces Table 1-2. Annunciator LEDs CW-90/90X Checkweigher...
Bar Graph LEDs The bar graph LEDs provide you with a fast way of determining if a container is too heavy ( ), too light OVER ), or is within an acceptable weight range ( UNDER ACCEPT ACCEPT OVER UNDER Slightly under Slightly over Moderately under...
Install feet onto bottom side of the column. 5. Turn the CW-90/CW-90X upright and replace the platter on the platform. 6. Attach indicator to the column with the two knobs and nylon washers provided. Position nylon washers between indicator enclosure and column mounting holes.
Making Power Connections The power source used for the CW-90 must be properly grounded to an acceptable earth ground. If the indicator is remotely mounted, the platform must be separately grounded from the chassis ground screw located on the bottom of the platform.
Port 1 supports full duplex RS-232 communications only; Port 2 provides either active 20 mA output or duplex RS-232 transmission. Both ports are configured using the SERIAL menu. See Section 3.4.4 on page 29. Connector Signal Port Ground Ground 20mA+ Ground Table 2-3. J2 and J3 pin assignments CW-90/90X Checkweigher...
1. Disconnect power to the indicator. Remove backplate as described in Section 2.4 on page 7. 2. Disconnect power supply cable from connector J9 on the CW-90/CW-90X CPU board. 3. Disconnect the wires at the following connectors: J1, J2, J3, J4.
5. Press down firmly to seat the option card in the CPU board connector. 6. Make connections to the option card as required. Use cable ties to secure loose cables inside the enclosure. When installation is complete, reassemble the enclosure as described in Section 2.8. CW-90/90X Checkweigher...
Page 15
Figure 2-5. CW-90/CW-90X CPU Board Jumper Description JP1/JP2 Jump excitation to sense. If using a 4-wire load cell cable, leave JP1 and JP2 on. If using a 6-wire load cell cable, take JP1 and JP2 off. Default is ON. JP3/JP4 Used when upgrading firmware.
During calibration, the ENTER key acts as a data entry confirmation key. It also acts as an EXECUTE key and accepts the value upon successful calibration. On the CW-90, the ZERO or CLR key can be used for a backspace.
Displays total calibration events. The display alternates (in one-second intervals) between CALIB and the four-digit number. CONFIG Displays total configuration events. The display alternates (in one-second intervals) between CFG and the Number four-digit number. EXIT Exits the audit trail and returns to weigh mode. Table 3-2. Audit Menu Parameters CW-90/90X Checkweigher...
The CW-90/CW-90X requires the WZERO and WSPAN points to be calibrated. The linearity points are optional, but must NOT duplicate zero or span. During calibration, the ENTER key acts as a data entry confirmation key. It also acts as an EXECUTE key, and accepts the value if calibration was successful.
Page 23
SCALE Menu Parameter Choices Description Level 3 submenus CALIBR WZERO See Section 3.3 on page 17 for descriptions, and see Section 4.0 on page 36 for WVAL calibration procedures. WSPAN Calibration can be performed in two places within the menu: the CALIBR menu shown in WLIN Figure 3-5 is an in-depth scale setup and calibration.
Page 24
(except for LFT parameters) now only needs to go below this display division value and then above it again. If checkweighing is used, THRESH should be less than the under value or it can disable the digital I/O points from tripping. Table 3-4. Scale Menu Parameters (Continued) CW-90/90X Checkweigher...
Page 25
SCALE Menu Parameter Choices Description Level 4 submenus PRIMRY UNITS Allows you to set the primary units, decimal point format, and display divisions. DECPNT DSPDIV Allows you to set the secondary units. Decimal point format and display divisions are selected automatically. Values are kg=kilogram (default);...
Page 26
Allows you to place the decimal point position. Use the ID ( ) and TARGET ( ) keys to 88888.8 place the decimal point where desired. 8888.88 888.888 888880 8.88888 88.8888 DSPDIV Display divisions. Selects the minimum division size for the primary unit’s displayed weight. Table 3-4. Scale Menu Parameters (Continued) CW-90/90X Checkweigher...
3.4.2 FEATUR Submenu AUDIT CALIBR SETUP TEST EXIT T&D ACCUM BRIGHT XXXXXXX Only displayed if Only displayed if turned on turned on from Setup menu from Setup menu CONFIG SCALE FEATUR SERIAL PFORMT DIGI/O VERS REMOTE CHKWGH REGION CONSC# 000000 CURVAL RESVAL MODE...
Page 28
Overriding passwords will clear configuration and calibration settings. To preserve settings (i.e., ID information), use Revolution software to upload ACCUM your data to a PC, then download it back to the CW-90 after the password override is performed. KEYLCK OVER...
Page 29
Accept band tolerance values. BOTH is a combination mode that allows the operator to have the CW-90/CW-90X acquire Accept band tolerances from actual weights on the scale, but then gives the operator an opportunity to digitally modify those values directly afterward....
Page 30
T&D Overriding passwords will clear configuration and calibration settings. To preserve settings (i.e., ID information), use Revolution software to your data ACCUM to a PC, then load it back to the CW-90 after the password override is performed. OVER UNLOCK...
3.4.3 FEATUR Menu, REGULA / INDUST Submenu AUDIT CALIBR SETUP TEST EXIT T&D ACCUM BRIGHT XXXXXXX Only displayed if Only displayed if turned on turned on from Setup menu from Setup menu CONFIG SCALE SERIAL PFORMT DIGI/O VERS FEATUR STNDBY RECALL CHKWGH CONSC#...
Page 32
The value specified for this parameter affects the function of the front panel TARE and ZERO keys. See Section 9.6 on page 64 for more information. Table 3-6. REGACY / INDUST Submenu Parameters (Continued) CW-90/90X Checkweigher...
3.4.4 Serial, PFORMT, DIG I/O, VERS Submenus AUDIT CALIBR SETUP TEST EXIT T&D ACCUM BRIGHT XXXXXXX Only displayed if Only displayed if turned on turned on from Setup menu from Setup menu CONFIG SCALE SERIAL PFORMT DIGIO VERS FEATUR DEFALT SOFTWR SCALE FEATUR...
Page 34
If any port is set to PRNMSG=ON, PRINT will display. TERMIN EOLDLY ECHO SOFTWR V 1.00 Displays the software version. LR,V.1.00 Displays the legally relevant version. DEFALT Resets settings to default values if YES is selected. Table 3-7. Serial, PFORMT, DIG I/O, VERS Submenu Parameters CW-90/90X Checkweigher...
Page 35
SETUP Menu Parameter Choices Description Level 4 submenus Turns the Digital input/output off. INPUT OVER Digital input functions. UNDER TARGET PRINT ZERO TARE UNITS ENTER CLEAR DSP ACC CLR ACC T&D DSP TAR CLR TAR KEY LCK HOLD CLR CN GRSNET OUTPUT FUNCTN...
Page 36
20-millisecond increments, maximum 65535) to turn the digital output off after the specified time (even though the function state has not changed). TRIP FRERUN FRERUN allows the output to toggle in motion; STABLE STABLE waits for standstill. Table 3-7. Serial, PFORMT, DIG I/O, VERS Submenu Parameters (Continued) CW-90/90X Checkweigher...
Test Menu AUDIT CALIBR SETUP TEST EXIT T&D ACCUM BRIGHT XXXXXXX Only displayed if Only displayed if turned on turned on from Setup menu from Setup menu DIG I/O COMM KEYPAD WARNING... PORT1 PORT2 PASS TEST DISCONNECT DIGITAL I/O FAIL PASS PASS BEFORE...
Setup menu 000000 CLEAR Figure 3-13. ACCUM Menu BRIGHT Menu There are eight brightness settings on the CW-90/CW-90X (0-7). If a number greater than 7 is selected, “RANGE” will be displayed because the number is out of range. AUDIT CALIBR SETUP...
ID Menu » » » » In order for the ID menu to appear, must be selected from the SETUP CONFIG FEATUR CHKWGH ID menu (see Figure 3-8 on page 23). AUDIT CALIBR SETUP TEST EXIT T&D ACCUM BRIGHT XXXXXXX Only displayed if Only displayed if turned on...
• Span calibration The CW-90/CW-90X requires the WZERO and WSPAN points to be calibrated. The linearity points are optional, but must NOT duplicate zero or span. During calibration, the ENTER key acts as a data entry confirmation key. It also acts as an EXECUTE key, and accepts the value if calibration was successful.
The Select Indicator dialog box appears. example, if the scale is calibrated with 1000 lb, a strain 3. Select CW-90 and click test may determine that at 2000 lb the calibration is 3 lb 4. From the Communications menu, select high.
Operation The CW-90/CW-90X can be configured for one of the three modes of operation: range, target weight, and target percent. Selecting which operation mode is done under the CHKWGH menu, MODE submenu (Figure 5-1). AUDIT CALIBR SETUP TEST EXIT T&D...
7. Using the keys to increment/decrement the values and the keys to move between digits, enter the desired under value. If using the CW-90, you can also use the numeric keypad to enter the desired under value. To discard any changes and return to weigh mode, press the Menu button.
12. Using the keys to increment/decrement the values and the keys to move between digits, enter the desired value. If using the CW-90, you can also use the numeric keypad to enter the desired under value. To discard any changes and return to weigh mode, press the Menu button.
2. Using the keys to increment/decrement the values and the keys to move between digits, enter the desired value. If using the CW-90, you can also use the numeric keypad to enter the desired target percent value. To discard any changes and return to weigh mode, press the Menu button.
(entering 5 equals 5%). If using the CW-90, you can also use the numeric keypad to enter the desired under value. ...
Page 48
TARE Enter keys to move between digits. If using the CW-90, you can also use the numeric keypad to enter the desired value. If a tare value is not being entered, press to move to the next parameter.
If an ID does not have values assigned, NO ID is displayed. The ZERO key acts as a backspace on the CW-90X. Use this key to navigate from a two-digit ID to a one-digit ID. On the CW-90, use the CLR or ZERO key.
You can use negative checkweighing if you have an item which you want to take weight away from. For example, if a product on the CW-90/90X weighs 20 pounds and you want to remove weight in 2 pound increments, with negative checkweighing enabled, the Over value could be set to -3 and the Under value could be set to -1.
Page 51
If using the CW-90, you can also use the numeric keypad to enter the desired under value. The Under value you are entering is a negative value. However, it will appear to be a lower number than the Over value because the negative symbol is not displayed.
Serial Commands The CW-90/CW-90X can be controlled by a PC or Command Function remote keyboard connected to an indicator serial port. KDISPACCUM Display ACCUM (pseudo key) Control is provided by a set of serial commands that KDISPTARE Display tare (pseudo key)
The command=?<ENTER> indicator must be in menu mode to use this function. VERSION Write CW-90/CW-90X software version Some parameters are valid only if other Write current displayed weight with units parameters or parameter values are specified.
Page 54
Table 6-4. SCALES Serial Commands Command Description Values EDP.BAUD#p Port baud rate 1200, 2400, 4800, 9600, 19200, 28800, 38400, 57600, 115200 EDP.BITS#p Port data bits/parity 8NONE, 7EVEN, 7ODD EDP.TERMIN#p Port termination character CR/LF, CR Table 6-5. SERIAL Port Serial Commands CW-90/90X Checkweigher...
Page 55
Command Description Values EDP.STOPBITS#p Port stop bits 2, 1 EDP.PRNMSG#P Print message ON, OFF EDP.ECHO#p Port echo ON, OFF EDP.EOLDLY#p Port end-of-line delay 0–255 (0.1-second intervals) EDP.TRIGGER#x Port trigger DEMAND, STREAM, COMAND, WAITSS, TARGET, L-STAB, OFF STR.POS Custom stream identifiers None, Space, + STR.NEG None, Space, +...
Page 56
CLRCN, GROSS, NET, PRIM, SEC, CLRTAR, CLRACC, TRIGGER DIO#b Digital output function OVER, UNDER, ACCEPT, ZEROBD DIO.TIMER#b Digital output timer function DIO.TRIP#x Digital output trip functionality STABLE FRERUN Digital inputs and outputs are specified by bit number Table 6-8. DIG I/O Serial Commands CW-90/90X Checkweigher...
6.1.6 Normal Mode Commands The normal mode print commands (see Table 6-9) transmit data to the serial port on demand in either setup or normal mode. Command Description Values CONSNUM Set consecutive number 0–9 999 999 Set unit ID nnnnnnn Set date MMDDYY, DDMMYY, YYMMDD, or YYDDMM.
Page 58
Configuration =00 (not used) =01 if primary DSPDIV=1 =10 if primary DSPDIV=2 =11 if primary DSPDIV=5 Configuration =00 (not used) =01 if secondary DSPDIV=1 =10 if secondary DSPDIV=2 =11 if secondary DSPDIV=5 Table 6-11. Custom Stream Format Identifiers CW-90/90X Checkweigher...
Page 59
Format Identifier Defined By Description Configuration =00 (not used) =01 if tertiary DSPDIV=1 =10 if tertiary DSPDIV=2 =11 if tertiary DSPDIV=5 Configuration =000 if current DECPNT=8888800 =001 if current DECPNT=8888880 =010 if current DECPNT=8888888 =011 if current DECPNT=888888.8 =100 if current DECPNT=88888.88 =101 if current DECPNT=8888.888 =110 if current DECPNT=888.8888 =111 if current DECPNT=88.88888...
Print Formatting The CW-90/CW-90X provides seven print formats, Command Description GFMT, NFMT, ACC.FMT, STRM.FM, BUS.FMT, HDRFMT1 . These determine the format of the HDRFMT2 <AD> Date of last accumulator event printed output when the key is pressed or when a PRINT <AT>...
7.2.1 Using the EDP Port The default CW-90/CW-90X print formats are shown in Table 7-2: With a personal computer, terminal, or remote keyboard attached to the CW-90/CW-90X EDP port, Format Default Format String Sample Output you can use the EDP command set to customize the print format strings.
Using Revolution, you can type text directly into the grid, then select weight value fields from the tool bar and place them where you want them to appear on the printed ticket. Figure 7-2 shows an example of the Revolution print formatting grid. Revolution Figure 7-2. Print Format Grid CW-90/90X Checkweigher...
On when not connected; off when connected Figure 8-1. WLAN board WLAN Card Installation Refer to CW-90/90X, 720i, 820i, iQUBE2, LaserLT WLAN Installation Instructions PN 108680 included with the WLAN option card for installation and configuration instructions. WLAN Option Card...
GRADS > 100,000 A value entered in configuration is out of the WVAL > 999999.0 acceptable range. Configuration value out of range Ranges for all parameters are listed in the Parameter Settings Commands section. Table 9-1. CW-90/CW-90X Error Messages CW-90/90X Checkweigher...
The XE EDP command can be used to remotely query Two EDP commands, P and ZZ, can be used to provide the CW-90/CW-90X for the error conditions shown on status about the indicator. These commands are the front panel. The XE command returns a 5-digit described in the following sections.
Continuous Output (Stream) Format Figure 9-1 shows the continuous output format sent to a CW-90/CW-90X port when that port’s TRIGGE parameter (SERIAL menu) is set to STREAM. <STX> <POL> <wwwwwww> <UNIT> <G/N> <S> <TERM> G = Gross ASCII 02 <CR> <LF>...
9.5.1 DIGFLx Parameters 3. Place the indicator in menu mode and set the digital filters (DFLTRx) to eliminate the The first three digital filtering parameters, DFLTR1, v i b r a t i o n e ff e c t s o n t h e s c a l e . ( L e a v e DFLTRL2, and DFLTR3, are configurable filter stages DFTHRH set to NONE.) Reconfigure as that control the effect of a single A/D reading on the...
Table 9-5. REGULAT / INDUST Mode Parameters, Comparison with Effective Values of Other Modes (Continued) Updating CW-90/90X Firmware To update firmware of the CW-90/CW-90X, you must have Revolution installed and a .hex file on your computer. Visit to download this free configuration software and the latest .hex file.
Page 70
Figure 9-5. Revolution CW-90/90X Screen. 7. The Rice Lake CW90 Updater screen appears. Specify the COM port the CW-90/CW-90X is connected to, change the baud rate if needed, and click the ellipses (...) to browse to and select the desired .hex file.
Specifications 115 VAC Capacity (enter in pounds (kg) or ounces) Column Est. Ship Part # Pounds Ounces Platform Dimensions Height Weight 105957 5 x 0.001 lb (2.5 x 0.0005 kg) 80 x 0.02 oz 10 x 10 x 4.25 in 12 in 29 lb 105958...
Page 72
1600 x 0.5 oz 12 x 12 x 4.25 in 12 in 33 lb Table 9-8. CW-90X Specifications American Style Plug used on 115 VAC and 230 VAC European Style Plug used on 230 VAC CE Marked/OIML Approved CW-90/90X Checkweigher...
Page 73
(20 mm) digits LED annunciators Center of zero, gross, net, tare, preset tare; percent, kg, g, lb, oz Keypad 21-key flat membrane panel (CW-90) Piezo membrane panel (CW-90X) Environmental Operating Temperature –10 to +40°C (legal); –10 to +50°C (industrial) Appendix...
CW-90/90X Limited Warranty Rice Lake Weighing Systems (RLWS) warrants that all RLWS equipment and systems properly installed by a Distributor or Original Equipment Manufacturer (OEM) will operate per written specifications as confirmed by the Distributor/OEM and accepted by RLWS. All systems and components are warranted against defects in materials and workmanship for two years.
1 to reach on-call personnel. Fax Number 715-234-6967 E-mail U.S. sales and product information at prodinfo@ricelake.com International (non-U.S.) sales and product information at intlsales@ricelake.com Mailing Address Rice Lake Weighing Systems 230 West Coleman Street Rice Lake, WI 54868 USA Appendix...
Need help?
Do you have a question about the CW-90 and is the answer not in the manual?
Questions and answers