Rice Lake SURVIVOR CW-80 Operating Manual page 12

Checkweigher
Hide thumbs Also See for SURVIVOR CW-80:

Advertisement

4.
Establishing ID Registers
NOTE:
In addition to the primary unit, up to three
alternate units (ALT 1, ALT 2, ALT 3) can
be established in the Setup mode. Refer to
p. 3-5 in the CW-80 Installation and
Service Manual for information on setting
up alternate units.
ID
Overview
When the Setup menu under "ID STR" is set to ON, up to 299
separate ID registers can be established. Each ID number is normally
associated with a product, and must have four categories of informa-
tion stored with it:
• Unit of measure
• Lower limit of ACCEPT band
• Upper limit of ACCEPT band
• Tare value
The operator can access these product ID registers by pressing the
ID key as shown on the separate CW-80 Operator's Card.
Setting Up an ID Register
Check that the desired units annunciator is lit. If a different unit is
active, press the UNITS key until the correct unit is lit. See note at
left.
Establish and store the desired tare.
Establish and store the desired lower ACCEPT tolerance (Un 1).
Establish and store the desired upper ACCEPT tolerance (Ov 1).
Press ID key to display the current ID number. Use the keys beneath
the digits to change the display to the ID number you want to
associate with the units, tare, and tolerances that you have just
established.
Immediately press the ID key a second time to store the completed
ID register. "STORED" will briefly be displayed.
If the 3-second time-out occurs before you press the ID key a second
time to store, the new values will not be stored with the displayed
ID. Instead, previous values stored with that ID number will be
recalled, and "RECALL" will be displayed. If this happens, begin
the procedure again.
Protecting ID Registers from Accidental Overwrite
To prevent an operator overwriting ID registers you have set up, you
can configure the CW-80 for an ID "read only" operation. The
operator can access any ID register by pressing the ID key but can't
change the values nor establish new registers.
Read-only operation must be established in Setup mode after you
have established all ID registers. Contact your installer or other
qualified scale technician for information on accessing the CW-80
Setup mode.
CW-80 Supervisor's Operating Guide
9