Dc Current; Capacitance - Fluke 28 II Ex Getting Started

True-rms digital multimeter
Hide thumbs Also See for 28 II Ex:
Table of Contents

Advertisement

28 II Ex
Getting Started

DC Current

Function
Range
600.0 μA
μA dc
6000 μA
60.00 mA
mA dc
[1]
400.0 mA
6.000 A
A dc
[2,3]
10.00 A
[1]
400 mA continuous; 600 mA for 18 hr maximum.
W 10 A continuous up to 35 °C. <20 minutes on, 5 minutes off at 35 °C to 55 °C. >10 A to 20 A for 30 seconds maximum, 5 minutes
[2]
off.
[3]
>10 A accuracy unspecified.

Capacitance

Range
10.00 nF
100.0 nF
1.000 μF
10.00 μF
100.0 μF
9999 μF
[1]
With a film capacitor or better, using the rel mode to zero residual.
30
Resolution
Burden Voltage
0.1 μA
100 μV/μA
1 μA
100 μV/μA
0.01 mA
1.8 mV/mA
0.1 mA
1.8 mV/mA
0.001 A
0.03 V/A
0.01 A
0.03 V/A
Resolution
0.01 nF
0.1 nF
0.001 μF
0.01 μF
0.1 μF
1 μF
MyFlukeStore
www.
Shop for Fluke products online at:
Accuracy
±(0.2 % + 4)
±(0.2 % + 2)
±(0.2 % + 4)
±(0.2 % + 2)
±(0.2 % + 4)
±(0.2 % + 2)
Accuracy
[1]
±(1.0 % + 2)
±(1.0 % + 2)
1.877.766.5412
.com