When To See An Authorised Pride Dealer For Service - Pride Go-Chair Owner's Manual

Hide thumbs Also See for Go-Chair:
Table of Contents

Advertisement

V I I .
C A R E
A N D
M A I N T E N A N C E

When to See An Authorised Pride Dealer for Service

The following symptoms could indicate a serious problem with your Go-Chair. If necessary, contact an authorised Pride
Dealer. When calling, have the model number, serial number, nature of the problem and the error code if available.
! Motor noise
! Frayed harnesses
! Cracked or broken connectors
! Uneven wear on any of the tyres
! Jerky motion
! Pulling to one side
! Bent or broken wheel assemblies
! Does not power up
! Powers up, but does not move
Corrective Maintenance
If the battery condition meter does not light up when you turn on the power:
! Check the harness connections. Make sure they are tight.
! Check the circuit breaker. Reset it if necessary.
! Check the battery connections.
If the above conditions prove normal, you can load test the batteries with a battery load tester. These testers are available
at automotive parts stores. Disconnect both batteries before load testing and follow the directions that come with the load
tester. If either one of the batteries fails the load test, replace both of them. If your Go-Chair still does not power up,
contact an authorised Pride Dealer.