Specifications - Nonin 9570 User Manual

Hide thumbs Also See for 9570:
Table of Contents

Advertisement

Available languages

Available languages

Specifications

Oxygen Saturation Display Range
Pulse Rate Display Range
Oxygen Saturation Declared Accuracy Range (A
Low Perfusion Oxygen Saturation Declared
Accuracy Range (A
Pulse Rate Declared Accuracy Range(A
Low Perfusion Pulse Rate Declared Accuracy
Range (A
*)
rms
Measurement Wavelengths and Output Power
Red
Infrared
*± 1 A
represents approximately 68% of measurements.
rms
*)
rms
0% to 100%
18 to 321 beats per minute
(BPM)
*)
70% to 100% SpO
rms
70% to 100% SpO
*)
20 to 250 BPM ± 3 digits
rms
40 to 240 BPM ± 3 digits
660 nanometers @ 0.8mW
Max Average
910 nanometers @ 1.2mW
Max Average
± 2 digits
2
± 2 digits
2
20

Advertisement

Table of Contents
loading

This manual is also suitable for:

9570 go2

Table of Contents