Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The Dexcom G4 System is intended for use in the electromagnetic environment
specified below. The customer or the user of the Dexcom G4 System should
ensure that it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
126
|
Dexcom G4 Continuous Glucose Monitoring System
Compliance
Electromagnetic Environment
Group 1
The Dexcom G4 System uses RF energy
only for its internal function. Therefore, its
RF emissions are very low and are not
likely to cause any interference in nearby
electronic equipment.
Class B
The Dexcom G4 System is suitable for
use in all establishments other than
domestic establishments and those
Not applicable
directly connected to the public low-
voltage power supply network that
supplies buildings used for domestic
Not applicable
purposes.
Guidance