Operational Test
Use the following procedure to test the basic
operability of your GMM test tool:
1. With no inputs connected, turn the GMM test tool
on by rotating the selector to VAC (
Meter display mode should appear, with a
random display in the mV range.
2. Rotate the selector to VDC (
primary display in the mV range.
3. Rotate the selector to Ohms (
display of "OL".
4. Connect the test leads to V and C.
5. Short the test lead tips and observe a display of
approximately 0.250 ohms.
6. Press 2 (/), short the test leads again, and
listen for a beep.
For a complete performance test of GMM test tool
functions and specifications, refer to the 863/867B
Service Manual (PN 689312).
v
). Combo or
E
) Observe a similar
R
). Observe a
Self Test
Begin an internal diagnostic routine by selecting SET
UP, wait two seconds, press d2. The display
now presents the following information:
1. A test pattern.
2. An error message (if a problem is encountered
during the test). The GMM test tool Model and
serial numbers.
3. The calibration status.
Generally, you cannot use the GMM test tool keys
during Self Test. (You can press 1 to exit the
diagnostic routine and return to the SET UP
Softkeys.)
User Maintenance
7