867B/863
Users Manual
m
WARNING
Hazardous voltages are present on
the External Trigger input terminal
and any attached probe when the
GMM is measuring voltages greater
than 30V rms or 60V dc.
Replacing Fuses
The following fuses should be used:
F 440 mA, 1000V, Minimum Interrupt Rating
10,000A.
F 11A, 1000V, Minimum Interrupt Rating
17,000A.
Warning
Use of fuses with voltage ratings
below 1000V reduces protection.
Do not replace the 440 mA 1000V
fuse with a 4/10A 600V fuse.
m
x
$
WARNING risk of electric shock.
m
CAUTION see explanation in manual.
s
Equipment protected throughout by DOUBLE
INSULATION or REINFORCED INSULA-
TION.
m
m
Overload Protection
Voltage or Current Input: 1000V
Surge Protection: 8 kV
Maximum Voltage Isolation to Ground: 1000V
Protection Levels: per IEC 1010-1; 1000V CAT III
(fixed distribution-level installations which can
experience peak voltage transients up to
8000V.)
s
Protection Class: II