Frequency
Gate Time .......................................................... 131 ms
Measurement Method ....................................... AC-coupled input using the ac voltage measurement function.
Settling Considerations ................................... When measuring frequency after a dc offset voltage change, errors
Measurement Considerations ......................... To minimize measurement errors, shield inputs from external noise
Range
Frequency
20 Hz – 2 kHz
2 kHz – 20 kHz
100 mV to
[1,2]
750 V
20 kHz – 200 kHz
200 kHz – 1 MHz
Notes:
[1]
Input > 100 mV
7
[2]
Limited to 8* 10
V Hz
Continuity
Continuity Threshold........................................ 20 Ω
Test Currents .................................................... 1 mA
Response Time ................................................ 100 samples/sec with audible tone
Rate.......................................................... Fast
Maxiumum Reading....................................199.99 Ω
Resolution............... ..................................0.01 Ω
Diode Test
Response Time ................................................. 100 samples/sec with audible tone
Rate..........................................................Fast
Maxiumum Reading....................................1.9999 V
Resolution.................... .............................0.1 mV
may occur. For the most accurate measurement, wait up to 1 second
to allow input blocking RC time constant to settle.
when measuring low voltage, low frequency signals.
Accuracy
Accuracy
90 days
23 °C ± 5 °C
0.01 + 0.002
0.01 + 0.002
0.01 + 0.002
0.01 + 0.004
Introduction and Specifications
Electrical Specifications
Temperature
Coefficient/°C
1 year
Outside 18 – 28 °C
23 °C ± 5 °C
0.01 + 0.003
0.002 + 0.001
0.01 + 0.003
0.002 + 0.001
0.01 + 0.003
0.002 + 0.001
0.01 + 0.006
0.002 + 0.002
1
1-13