STEP 2. Apply control solution
Shake the control solution vial thoroughly before use. Squeeze out a drop
and wipe it off, then squeeze out another drop and place it on the tip of
the vial cap. Hold the meter to move the absorbent hole of the test strip to
touch the drop. Once the confirmation window fills completely, the meter
will begin counting down. To avoid contaminating the control solution, do
not directly apply control solution onto a strip.
STEP 3. Read and compare the results
After the meter counts to 0, the result of the control solution test will
appear. Compare the result with the range printed on the test strip vial.
The result should fall within this range.
Out-of-range results
If you continue to have test results fall outside the range printed on the
test strip vial, the meter and strips may not be working properly. Do NOT
test your blood. Contact the local customer service or place of purchase
for help.

TESTING YOUR BLOOD

Preparing the Lancing Device for Blood Testing
Please follow the instructions in the lancing device insert for collecting a
blood sample.
STEP 1. Insert the test strip to turn on the meter
Wait for the meter to display " ".
STEP 2. Select the appropriate measuring mode by pressing Main
button
For selecting the measurement mode, please refer to the FOUR
MEASURING MODES .