What You See
42
What It Means
Meter has detected an
error; accurate test result
not possible.
Meter has detected an
error; accurate test result
not possible.
There was a problem filling
the test strip completely
full.
Your Blood Glucose Level is
very high or meter error.
What You Do
• You may have applied
blood too early to the test
strip. Repeat test and wait
until you see the flashing
blood drop to apply blood.
• There might also be
moisture on the test strip
guide. Clean meter if
needed and dry
thoroughly (see pg. 27).
Repeat test.
• Test strip is not
completely filled. Release
used test strip and repeat
test. See pg. 8, or
• 10-test disc may have been
damaged due to intense
heat or moisture exposure.
Replace 10-test disc, using
a new box if necessary.
Review pg. 8, and repeat
test.
Turn meter off, then on
using
. Retest. If error
appears again, call your
physician or healthcare
professional (HCP). Your
blood glucose may be
above 1500 mg/dL (83.3
mmol/L). If the problem is
with the meter and cannot
be corrected, contact your
Bayer Customer Service
Department (see pg. 45).