Normal Control Solution
Strip
Lot
Operator
A
Lay User (n=50)
B
Lay User (n=50)
High Control Solution
Strip
Lot
Operator
A
Lay User (n=50)
B
Lay User (n=50)
Accuracy
1. Bayer Accuracy Evaluation:
To assess the accuracy of the Ascensia B
fresh fingerstick specimens from people with diabetes were
tested with the Ascensia B
and the YSI 2300 STAT Plus Glucose Analyzer. The glucose
concentration of the samples ranged from 42 to 542 mg/dL
with an average of 178 mg/dL. The range of hematocrits
was 29% to 55% with an average of 40%.
No. of
Mean
Assays
(mg/dL)
100
151.1
100
149.7
No. of
Mean
Assays
(mg/dL)
100
336.7
100
339.6
(2 readings per fingerstick)
REEZE
Total
SD
(mg/dL)
7.2
6.4
Total
SD
(mg/dL)
17.3
14.4
system, 102
REEZE
Total
CV
(%)
4.7
4.3
Total
CV
(%)
5.1
4.2
33