NPB-40 Operation
Press the Menu button while in normal mode to access
the parameter set up menu hierarchy. Press the Menu
button seven times to access the print data display.
The summary report is displayed if the NPB-40 is
currently posting saturation and rate. It then lists
snap-shot data stored since the NPB-40 is powered on,
and any sensor trend data.
If snap-shot data are stored in the NPB-40 memory, the
Data icon is displayed on the print data display. If
sensor-event data are stored in sensor memory, the Data
In-Sensor icon is displayed on the print data display. If
both types of data are stored, both icons are displayed. If
no data are stored, neither icon is shown (blank display).
Note: The monitor and sensor does not differentiate
between patients; therefore, use caution when
reviewing the report as the list of snap-shots may
contain data from more than one patient. Also, if a
single-use sensor is used more than once, the
sensor trend report section can contain data from
more than one patient.
Note: The snap-shot list is deleted when the NPB-40 is
turned on. When first connecting a single-use
sensor, look for the DATA in SENSOR icon.

Procedure

With the NPB-40 in the normal mode: