Set Alarm Limits
Discussion
The Alarm Limit display allows the user to adjust the
upper and lower saturation and pulse rate limits.
Pressing the Up Arrow or Down Arrow button for three
seconds or more causes the increment or decrement to
scroll.
The Alarm Limit Changed indicator is displayed
anytime an alarm limit is being changed or has been
changed. See Alarm Limit Changed Indicator on page 63.
Note: When the NPB-40 times-out (30 seconds), the
currently shown parameter is set and the NPB-40
display returns to the normal mode.
Procedure
With the NPB-40 in the normal mode:
1. Press the Menu button. The SpO
is shown.
NPB-40
NPB-40 Operation
lower alarm limit
2
59