List of Error Codes

Error code
Inflation error
Deflation error
Overpressure error
Insufficient inflation error
Indeterminable blood pressure error
Low pulse level error
Blood pressure error
Pulse rate error
Device error
28
Explanation
• When the pressure does not exceed 12
mmHg within the set time after the start of
inflation
• When the inflation does not reach the set
cuff pressure within the specified time
after the start of inflation
• When the deflation speed is too fast
during the measurement
• When the deflation speed is too slow
during the measurement
• When the measurement does not finish
within the specified time after starting the
measurement
• The cuff pressure exceeded 299 mmHg.
• Blood pressure could not be measured
due to insufficient inflation level.
• Blood pressure could not be measured
even when the cuff pressure reached the
specified pressure.
• Pulse wave was too small.
• Relationship between systolic and
diastolic pressures was abnormal.
• Pulse rate did not stay within the range of
30 to 199 beats/min.
• This is a trouble of the main unit of the
monitor.
How to correct
• Confirm that the air tube connecting the cuff and
the main unit is connected securely.
• Confirm that the air flow in the air tube
connecting the cuff and the main unit isn't being
restricted.
• Confirm that the cuff is wrapped correctly (refer
to pages 14 and 15).
• Check bladder for leaks and, if necessary,
replace the bladder with new one (option).
• Confirm that air flow in the air tube connecting
the cuff and the main unit isn't being restricted.
• If the measurement is made by setting the P-
SET to "AUTO", ask the patient not to move
during the inflation.
• Confirm that the P-SET is securely set to
"AUTO". Turn the Knob counterclockwise as far
as it goes until you can hear a click sound.
• If the measurement is made by manual inflation
level setting, set the value to 30 to 40 mmHg
higher.
• Confirm that the cuff is wrapped correctly (refer
to pages 14 and 15).
• Confirm that the cuff is wrapped correctly (refer
to pages 14 and 15).
• Ask the patient not to move during the
measurement.
• Check the patient for arrhythmia.
• Contact Omron's repair department toll-free at
(1-877-216-1336).