Your Setra Quick Count scale is highly stable, and requires infrequent
calibration. While it may be span calibrated as often as desired, the
factory calibration is adequate for most counting applications. For
precise
weighing
desirable. To perform a span calibration on Quick Count scales with
operating program version 1.3 or higher, use the following procedure:
1. Remove all objects from the pan.
2. Make certain the scale is level.
3. Press the units key to place the scale in either the GRAMS or POUNDS
weighing mode. (If either of these units of measure are disabled,
refer to " Units Enable/Disable Selection"
Technician's Guide.)
4. Press the zero key and while the scale is displaying " bUSY" *, press
the units key.
The scale will flash " CAL" and return to displaying " bUSY" .
5. The scale will then display " L xxxxx" , where xxxxx is the scale's
first calibration point. Press the sample size key repeatedly, the
scale will display each calibration point available.
To ensure maximum accuracy the calibration point selected should be
as close as possible to the scale's full capacity.
6. When the desired calibration point is displayed, load the displayed
weight on the pan, and press the units key.
7. The scale will display " bUSY" , and then return to the normal
weighing mode.
8. Remove the weights from the pan, press zero to zero the display.
*
If you are having difficulty entering the calibration mode, gently
touch the edge of the weighing pan to prolong the " bUSY"
message in step 4.
Setra Quick Count™ Operator's Manual
applications
more
Calibration
periodic
calibrations
in the Setra Quick Count
Appendix II
may
be
display
15