7. Immediately touch the tip of the test strip to the drop
of control solution. The solution is pulled into the test
strip through the tip.
8. Hold it in the drop until the meter beeps.
NOTE: The
does not indicate the control test
result is within the specified range. It only marks it
as a control solution test. It will not be included in
your averages.
®
22
You will see the meter
count down 5 seconds until
the test is complete and
your control test result is
displayed.
The meter will automatically
recognize and
mark the
control result for you.
9. Compare your control test result
to the ranges provided on the test
strip bottle or carton.
If your control test result is out of range, do not
use your meter for blood glucose testing until you
resolve the issue. See "What You Should Do" on
page 45.
To turn your meter off,
simply remove the test strip.
Dispose of the used test
strip as medical waste.
The result will be
automatically
stored in the meter memory.
Control results will not be
included in your blood
glucose averages or in the
7 day HI-LO summary.
IMPORTANT: To ensure accuracy of your blood
glucose test results and accuracy of your control
test results, use only C
ONTOUR
solutions.