b. Calendar Settings
After the printer setting has been determined in the setup mode, "y"
will appear in the SpO
display indicating calendar setup mode for the
2
year. The year may be set to "00" through "99". After selecting the
year, the display will show "nn" indicating the setup mode for the
month. The month may be set to "00" through "12". After selecting the
month, the display will show "d" indicating the setup mode for the day
of the month. The day may be set to "01" through "31". Refer to Figure
VI for a flowchart of setting the calendar. When the calendar setting
sequence has been completed, the 920 continues to the clock
settings (refer to section c. of this section).
NOTE:
Setting the month to "00" disables the clock
function and helps conserve battery life.
SpO
2
Display
Setup
Default
Setting
Mode
y
97
Press
nn
00
Press
d
01
Press
Proceed to
h
clock setup
Figure VI: Flowchart for setting the calendar.
13
Display
Press
Press
Increment
year
Press
Press
Press
Increment
month
Press
Press
Press
Increment
day
Press