SECTION 8—MAINTENANCE
Top Mounting Screw
Control Box
Bottom Mounting Screw
Bottom Washer
NOTE: The two plastic caps covering the top locknut and the top mounting screw are not shown.
*NOTE: Actuator power cord shown disconnected from the control box.

Maintaining the Base Adjustment

This procedure MUST be performed by a qualified technician.
NOTE: For this procedure, refer to  
FIGURE 8.2.
The base adjustment should not require 
any attention other than:
1. Check the squareness of the legs when 
in the closed position.
2. Place a 90° square on the inside of the 
legs and base to determine the 90° 
alignment.
3. Adjust the linkage rods until 90° 
alignment is achieved.
Reliant™ 440
Mast Mounting Bracket
FIGURE 8.1 Replacing Electric Actuator
WARNING
34
Lift Arm
Mounting
Bracket
Top Locknut
Top Washers
Electric Actuator
Actuator Power Cord*
Bottom Locknut
Bottom Washer
FIGURE 8.2 Maintaining the Base
Adjustment
Lift Arm
90°
Square
 
Linkage
Rod
Part No 1141431