I.V. ROD TELESCOPING KIT
9000XT RECLINER - KIT NO. 1074656
2000, 4000 AND 9000 RIDELITE
RECLINER - KIT NO. 1029598
NOTE: Check ALL parts for shipping damage. In case of
shipping damage, DO NOT use. Contact Carrier/Dealer
for further instruction.

SAFETY SUMMARY

The following recommendations are made for safe instal-
lation and adjustment of the I.V. Rod Telescoping Kit:
GENERAL WARNINGS
DO NOT install or use this equipment with-
out first reading and understanding this
instruction sheet. If you are unable to un-
derstand the Warnings, Cautions or Instruc-
tions, contact a healthcare professional,
dealer or technical personnel before at-
tempting to install this equipment - other-
wise, injury or damage may occur.
DO NOT pull/push the wheelchair by the
I.V. rod assembly. Injury to the assistant/
occupant and/or damage to the I.V. rod
assembly may occur.
DO NOT place any other objects on the
I.V. rod assembly other than those deter-
mined by a healthcare professional.
INSTALLATION WARNINGS
After ANY adjustments, repair or service
and BEFORE use, make sure all attaching
hardware is securely tightened - other-
wise injury or damage may result.
Use EXTREME caution when the I.V. rod
holder for MODELS 2000, 4000 and 9000
Ridelite is installed. When installed, the rear
arm release lever is removed and will NOT
lock the arm in place.
NOTE: If installing the telescoping I.V. rod holder onto
a 2000, 4000 or 9000 RIDELITE, refer to INSTALL-
ING THE TELESCOPING I.V. ROD HOLDER - 2000,
4000 or 9000 RIDELITE in this instruction sheet.
INSTALLING THE TELESCOPING
I.V. ROD HOLDER - 9000XT ONLY
(FIGURE 1)
NOTE: Refer to INSTALLATION WARNINGS in the
SAFETY SUMMARY in this instruction sheet.
Installation, Assembly and Operating Instructions
1. Remove the rear wheel from the wheelchair. Refer to
REMOVING/INSTALLING THE REAR WHEELS
in this instruction sheet.
2. Remove the arm from the wheelchair. Refer to RE-
MOVING/INSTALLING THE ARMRESTS in this in-
struction sheet.
3. Remove the locknut and hex screw that secure the
arm socket to the wheelchair frame.
4. Remove the locknut that secures the back tube to the
clevis bracket.
5. Position the I.V. holder on the existing pivot bolt as
shown in FIGURE 1.
Arm Socket
(OUTSIDE of
Wheelchair
Frame)
Locknut
Outside of
Wheelchair
Frame
Locknut
(STEP 8,9,11)
Side Frame
I.V. Holder
(STEP 5)
Mounting
Screw
(STEP 7)
Front
Cylinder
FIGURE 1 - INSTALLING THE
TELESCOPING I.V. ROD HOLDER
1
Hex
Screw
STEP 3
Back Tube
(STEP 4)
Clevis
Bracket
(STEP 4)
Pivot Bolt
(STEP 5)
Flat Portion
(STEP 5)
Locknut
(STEPS 4,6,11 )

Advertisement

loading

Summary of Contents for Invacare 2000

  • Page 1: Safety Summary

    NOT lock the arm in place. NOTE: If installing the telescoping I.V. rod holder onto a 2000, 4000 or 9000 RIDELITE, refer to INSTALL- ING THE TELESCOPING I.V. ROD HOLDER - 2000, 4000 or 9000 RIDELITE in this instruction sheet.
  • Page 2 6. Reinstall locknut onto pivot bolt. DO NOT tighten. 7. Install the provided mounting screw through the lower hole on the front cylinder section of the I.V. holder and into the side frame. 8. Install locknut onto mounting screw. 9. Turn locknut until snug fit is obtained. DO NOT tighten. 10.
  • Page 3 INSTALLING. 1. Secure rear wheel, patient operated wheel lock, and spacers to wheelchair frame with axle mounting screw and locknut. Securely tighten. Wheelchair Frame Locknut Patient Operated Spacers Rear Wheel FIGURE 3 - REMOVING/INSTALLING THE REAR WHEELS - RECLINERS WITH PATIENT OPERATED WHEEL LOCKS Recliners With Attendant Operated Wheel Locks (FIGURE 4)