Setting Parameters; Numeric Parameters - Larson Davis 820 Technical Reference Manual

Sound level meter
Hide thumbs Also See for 820:
Table of Contents

Advertisement

Where n is from 1 to 8, indicating the group list index and
var_no is the number of the "R" variable associated with the
index. Use var_no of "0" to define the last group list index
when n<8.
Example
G1, 4 cr
assigns the current SPL to the first group index
G2, 15 cr
assigns Lmin to the second index
G3, 19 cr
assigns Lmax to the third index
G4, 0 cr
terminates group command programming
After programming this group read list, the response to G0 is
59.5, 38.6, 102.2, the SPL, L

Setting Parameters

All commands begin with S. See Chapter 8 for a list of all the
parameters and their settings. Note that the parameter numbers
are for both Set (S) and Query (Q) commands. There are four
types of parameters:
Numeric
Indexed
Character strings
Template
Brackets indicate optional characters and operands.
cr = carriage return; lf = line feed; _ = space

Numeric Parameters

Syntax
S[et]item_number, parameter_valuecr
B-12
and L
.
min
max
Response
lf
Model 820 Technical Reference Manual