Larson Davis 820 Technical Reference Manual page 94

Sound level meter
Hide thumbs Also See for 820:
Table of Contents

Advertisement

7
Current Date
8
Day of Week
9
Baud Rate
10
RS-232 Address
12
COM Ports Hand-
shaking Register
8-8
(mm/dd/yy)
[Sun-Sat]
[300-19200]
(0-127)
[None, Hdwr, XOFF,
Both]
Leap years are considered. For years
less than the year manufactured, the
21st century is assumed (2000-2092),
otherwise the 20th century is used
(1993-1999)
Baud rate of the serial interface.
Serial printer or computer rate should
be the same.
Enables addressed communication
mode for a network of Larson Davis
SLMs. Each must have a unique
address. Normally only one Model
820 is used, thus the address should
be set to 0 to disable.
Serial handshaking protocol. Model
820 -SW1 uses [None]. Note: To con-
serve power, the Model 820 turns off
the serial port after 255 sec. of non-
use. To keep unit on and connected,
an external power supply must be
used.
Model 820 Technical Reference Manual