Chapter 6 Troubleshooting - Omron HGM-112 Instruction Manual

Table of Contents

Advertisement

BLOOD GLUCOSE MONITORING SYSTEM-HGM-112

Chapter 6 Troubleshooting

When the meter is used incorrectly, in poor condition, or other unexpected
events happen, the screen will show the error message to remind the user
should notice the problem or show how to resolve them.
The following messages may mean you have obtained a result that
requires immediate attention or there may be a problem with the meter.
Message
HI
LO
Temperature
Error
Low battery
warning
Wrong Test
Strip
Wrong Blood
Enter Timing
18
Symbol
Mean
You may have a very high
blood glucose level.
( > 600 mg/dL)
You may have a very low
blood glucose level.
(< 20 mg/dL)
You may have tested near
the low or high end of the
meter s operating
temperature range.
(10 C ~ 40 C/50 F ~ 104 F)
The power is too low to
perform a test.
Test strip could be expired,
upside down,broken, or
used again.
Obtain blood sample too
soon.
Action
1. Do control solution test.
2. Re-do your glucose test.
3. Contact your healthcare
professionals.
1. Do control solution test.
2. Re-do your glucose test.
3. Contact your healthcare
professionals.
Repeat the test after the meter
has reached a temperature
within the operating range.
Replace the battery.
1.make sure if strip inserting
correctly.
2.use an New HGM-STP1A
test strip if broken, used.
Follow instruction step.