BLOOD GLUCOSE MONITORING SYSTEM-HGM-112
Precision:
This study shows the variability from strip to strip in sample tests. The
results are shown in the following table.
Within Run Precision:
Samples
Sample 1
Sample 2
Sample 3
Sample 4
Sample 5
Between Day Precision:
Control solution 1
Control solution 2
Control solution 3

Chapter 8 Specifications

8.1 Specification of HGM-112 meter

Measuring range: 20~600mg/dL
Calibration: Plasma-calibration
Measuring unit: mg/dL
Measuring time: 5 seconds
Sample type: Capillary whole blood samples drawn from the fingertips
or palm
Blood Sample size: 1μL
Operating temperature: (10 C ~ 40 C) / (50 ºF ~ 104ºF)
Relative operating humidity: Below 90%RH
Memory capacity: one memory for the lastest memory
Dimension: 63.2 mm x 40.6 mm x 18.0 mm (L×W×H)
Weight: 25.5 g with battery
Power Supply: one CR2032 battery
Storage temperature: -20 C ~ 50 C (-4 F ~ 122 F)
Please protect the HGM-112 from impact
Please do not disassemble the HGM-112
20
Mean glucose concentration
41.01 mg/dL
83.49 mg/dL
116.82mg/dL
177.52mg/dL
177.52mg/dL (9.85 mmol/L)
341.61mg/dL
177.52mg/dL (9.85 mmol/L)
42.83 mg/dL
134.36 mg/dL
334.93 mg/dL
measured CV
SD= 4.02 mg/dL
4.26%
4.85%
4.66%
4.05%
3.61 mg/dL
4.20%
3.44%