Chapter 6 Instructions on Caring and Storing for Your Strips and Meter 6.1 Caring for the strip 6.2 Caring for the meter Chapter 7 Troubleshooting Chapter 8 HEA-232 Performance Characteristics Chapter 9 Specification OMRON blood glucose monitoring system HEA-232.indd 1 2012/6/13 下午 06:44:18...
11. The device for self-testing is used for the monitoring of an existing disease (eg., diabetes) the patient should only adapt the treatment if he has received the appropriate training to do so. OMRON blood glucose monitoring system HEA-232.indd 2 2012/6/13 下午 06:44:18...
The device should not be used for screening or diagnosis of diabetes or for testing on neonates. Intended Use The HEA-232 is an in-vitro diagnostic medical device intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips. Testing is done outside the body.
HEA-232 utilizes electrochemical biosensing technology (Amperometric biosensor) to measure the glucose level in fresh capillary whole blood. At the applied potential after blood draw into the reaction zone of strips, conversion of electroactive species generated in the enzyme layer occurs at the electrode and the resulting current is measured by HEA-232 .
Page 7
Inserting Bar- Insert this test strip port with this side facing up. C. Function Display Test Unit Battery Temperature Test Result Error Message mg/dL mmol/L Battery Test Unit Temperature Error Message Test Result OMRON blood glucose monitoring system HEA-232.indd 5 2012/6/13 下午 06:44:19...
3. Insert battery cover back to the battery compartment until a “click”sound is heard. 4. Turn on the device to see whether the HEA-232 is powered on. If not, please re-examine battery replacing procedures. 5. Please dispose of used batteries according to local regulations.
Remove the cap by twisting it b. Install the lancet d. Remove the cap if lancing c. Twist the protective cap of lancet device and adjust the puncture depth OMRON blood glucose monitoring system HEA-232.indd 7 2012/6/13 下午 06:44:23...
Page 10
Take a new blood glucose test strip from the vial and insert it into the meter correctly as the following. Insert the test strip (Correctly) Do not insert the strip from reverse side OMRON blood glucose monitoring system HEA-232.indd 8 2012/6/13 下午 06:44:23...
Page 11
The meter will automatically start and show the last result data; if no data, it will show the following. mg/dL Caution:The HEA-232 meter measurement unit was set as mg/dL or mmol/L from manufacturer side. the unit exchange ratio is as the followings: 1 mmol / L = 18.02 mg/dL 2.
Page 12
The HEA-232 will display the code number on the screen w h e n w a i t i n g f o r t h e b l o o d t o d r o p i n , t h e u s e r n e e d s t o make sure the code number is the same as that of the strip container.
Page 13
Caution:The HEA-232 measurement unit was set as mg/dL or mmol/L from manufacturer side. Caution: Do not smear blood on the strip. Touch the blood drop of fingertip to the reaction zone. Sample filled Sample too small The following items will cause inaccurate results: □...
You think your glucose meter could be brok en. If you will like to perform control solution test or have any question, please contact your local authorized distributor for more information. OMRON blood glucose monitoring system HEA-232.indd 12 2012/6/13 下午 06:44:26...
Page 15
Touch the tip of the test strip to the drop of control solution. The control solution test result will show on display after 5 seconds count down. The HEA-232 will automatically power off after the test strip is removed. Discard the used test strip in accordance with local regulations.
Chapter 5 Memory Function Reviewing past results ◆ Press button to turn on the HEA-232, the display will show the most recent test result then the index and then the most recent test result after a 1 second delay. Press the button again to display the second recent test result; every time the button is pressed, it will display a previous result.
Page 17
After 1 second After 2 seconds Caution:The HEA-232 meter measurement unit was set as mg/dL or mmol/L from manufacturer side. the unit exchange ratio is as the followings: 1 mmol / L = 18.02 mg/dL OMRON blood glucose monitoring system HEA-232.indd 15...
Do not allow any liquid to enter the test strip slot or the button of the meter. • Do not put the meter under water or any liquid. • The meter must be stored at room temperature • OMRON blood glucose monitoring system HEA-232.indd 16 2012/6/13 下午 06:44:27...
W r o n g Te s t Use other new HEA- expiration date, broken, Strip STP30 Test Strip. or used again. Wrong Blood Obtain blood sample too Follow instruction step. Enter Timing soon. OMRON blood glucose monitoring system HEA-232.indd 17 2012/6/13 下午 06:44:28...
Chapter 8 HEA-232 Performance Characteristics Traceability: The HEA-232 test result were calibrated to YSI readings. The YSI analyzer was calibrated using a standard solution(180mg/ dL(10mmol/L) which can be traced to NIST SRM #917A. Accuracy: The results from the HEA-232 were compared with those from YSI glucose analyzer.
The measurement unit either mg/dL or mmol/L is fixed in the meter according to the country demand. You are not able to change the unit in the meter. Caution: Any further service and product information, please contact OMRON distributors. OMRON blood glucose monitoring system HEA-232.indd 19 2012/6/13 下午 06:44:28...
Page 22
DO NOT USE IF PACKAGE IS DAMAGED For self-testing DELBio, INC. 3F, 252 Shangying Road, Guishan Industrial Zone, Taoyuan County 33341, Taiwan www.delbio.com.tw Deltronics (Netherlands) BV Zandsteen 15, 2132 MZ Hoofddorp, The Netherlands 0120 OMRON blood glucose monitoring system HEA-232.indd 20 2012/6/13 下午 06:44:28...