OMRON blood glucose monitoring system HEA-232.indd 17

Chapter 7 Troubleshooting

When the meter is used incorrectly, or other unexpected events occur,
the screen will show an error message to remind the user to notice the
problems and/or how to resolve them.
The following messages mean you may have obtained a result that
requires immediate attention or there may be a problem with the meter:
Message
Symbol
HI
LO
Temperature
symbol
continuously
flash
Battery symbol
continuously
flash
W r o n g Te s t
Strip
Wrong Blood
Enter Timing
Mean
Although this message
could be due to a test
Blood glucose may be
error, it is safer to treat first
higher than the measuring
and then do another test.
range of the system
1. Do control solution test.
( > 6 0 0 m g / d L ; 3 3 . 3
2. Redo your glucose test.
mmol/L )
3. Contact your healthcare
Although this message
could be due to a test
Blood glucose may be
error, it is safer to treat first
lower than the measuring
and then do another test.
range of the system
1. Do control solution test.
(< 20 mg/dL; 1.1 mmol/L)
2. Redo your glucose test.
3. Contact your healthcare
You may have tested
in an environment near
Repeat the test after the
the low or high end of
m e t e r h a s r e a c h e d a
the system's operating
temperature within the
temperature range.
operating range.
(10℃ -40℃ /50 ℉ -
104 ℉ )
The power is too low to
Replace the battery.
perform a test.
Test strip is over
Use other new HEA-
expiration date, broken,
STP30 Test Strip.
or used again.
Obtain blood sample too
Follow instruction step.
soon.
Action
professionals.
professionals.
17
2012/6/13 下午 06:44:28