When to run the glucose control solution test
□ The test strip vial is opened for long time or you think the test strips
have been damaged.
□ When the meter could be damage or not working properly.
□ When you want to double check on the meter's performance.
Running a Control Test:
1. Preparing a Test Strip
1) Insert the test strip into the meter. The meter will display the last
measurement data.
2) Then the screen shows the drop blood symbol.
2. Running the Glucose Control Solution Test
3) Before using the control solution, gently shake the control bottle to mix
well.
4) Remove the control bottle cap and wipe the tip of the bottle with a
tissue.
5) Squeeze the bottle until a tiny drop forms at the tip of the bottle and
put this drop on a clean surface.
6) Touch the drop to the tip of the confirmation window of the test strip.
When the confirmation window is filled with control solution
completely, the meter will automatically start to run the measurement.
7) After 5 seconds, the screen will display test result. The meter will
automatically turn off after the test strip is removed. Discard the used
test strip in accordance with local regulations.
* Write the date you first open the bottle on the bottle label. The control
solution must be discarded after three months from the date the bottle
was opened, on the "discard date or on the "expired date on the
bottle label, whichever comes first. Do not use control solution that is
past the "expired or discard date.
Warning: Before running test glucose control solution test, always
check the expiration date. DO NOT use control solution if
expired over three months of first opened day.