ILX Lightwave LDT-5500B Series User Manual page 46

Table of Contents

Advertisement

4
C H A P T E R
TEC:T
TEC:T?
Parameters
Notes
Examples
TEMP_LIMIT?
Parameters
Notes
Examples
MODEL?
Response
Examples
32
LDT-5500B Series
C O M M A N D R E F E R E N C E
LDT-5500B Series Device-Dependent Commands
The TEC:T command sets the TECs constant temperature set point.
The TEC:T? query returns the value of the TEC temperature measurement.
A value which represents the temperature set point, in
The TEC temperature is controlled to this set point only when the TEC is operated in T
mode. If the maximum temperature limit is set (see TEC:LIM:T), this limit becomes the
maximum set point value as well.
"TEC:T 15.0" - action: sets the TEC temperature set point to 15.0
"TEC:T?" - response: 29.2, means the measured load temperature is 29.2
The TEMP_LIMIT? query returns the current status of the temperature limit error.
None.
This error is always updated. This means if the output was disabled because of a
temperature limit, the user may not know this if this query is perfomed AFTER the actual
temperature is below the temperature limit setpoint.
"TEMP_LIMIT?" - response: 0, means that the current measured temperature is below
the set temperature limit.
"TEMP_LIMIT?" - response: 1, means that the current measured temperature is above
the set temperature limit.
The MODEL? query requests the instrument to identify itself by model number.
Returns an ASCII identification string of the model number.
"MODEL?" - response: 5545B, means the instrument is an LDT-5545B.