LDT-5500B Series
C O M M A N D R E F E R E N C E
LDT-5500B Series Device-Dependent Commands
The TEC:C3 command sets the TECs Steinhart-Hart equation constant C3.
The TEC:C3? query returns the value of the Steinhart-Hart equation constant C3.
Steinhart-Hart value between the value of 9.999 and -9.99.
When the LM335 or AD590 sensors are selected via the SENSOR SELECT switch, only
C1 and C2 are used.
"TEC:C3 1.125" - action: sets the Steinhart-Hart constant C3 for the current selected
sensor to 1.125.
"TEC:C3?" - response: 2.100, means the Steinhart-Hart constant C3 is set to 2.100.
The TEC:CAL:ITE command is used to enter into remote TEC current user calibration
mode.
None.
Calibration mode can be disabled by sending the output on command or pressing the
output on button.
"TEC:CAL:ITE" - action: sets the unit into remote TEC current calibration.
The TEC:CAL:SEN command is used to enter into remote SENSOR user calibration
mode.
None.
Calibration mode can be disabled by sending the output on command or pressing the
output on button. The calibration mode is dependent upon which setting the SENSOR
SELECT switch is set to.
"TEC:CAL:SEN", and SENSOR SELECT switch set to AD590 - action: enters the unit
into remote calibration mode for AD590.