INSTALLATION

1. Unpack scale and infant tray and note any discrepancies in its condition.
Keep carton and packing for return shipment of instrument if necessary.
2. Place scale on a stable, vibration-free location away from direct sunlight
and away from any rapidly moving air source. Make certain power cord is
stored out of the way of normal traffic.
3. Before mounting infant tray onto scale, if necessary, level scale by
adjusting the four legs on bottom of scale, up or down, until the bubble in the
level gage, located on the weigh bridge, is centered as shown in illustration
below. After level is achieved, lock legs in place by tightening lock nuts on
legs against scale bottom.
Adjusting
Legs
4. Place infant tray on scale with inch measure on tray facing operator. Align
screw holes and secure tray to weigh bridge using the two nylon screws
provided. Do not overtighten the nylon screws. Make certain infant tray is not
touching any surrounding material or objects.
Do not place scale on any unstable cart, stand, or
table. The scale may fall, causing injury to an infant
or adult and seriously damage the unit; or proper
operation of the scale may be inhibited.
3
Leveled
Not Leveled