Measurement With The Desired Systolic Pressure - A&D UA-767NFC Instruction Manual

Digital blood pressure monitor
Hide thumbs Also See for UA-767NFC:
Table of Contents

Advertisement

Measurement with the Desired Systolic Pressure

Model UA-767NFC is designed to detect the pulse and to inflate the cuff to a
systolic pressure level automatically.
Use this method when re-inflation occurs repeatedly
or when the results are not displayed even if the
pressure decreases to 20 mmHg or less.
When the memory setting is 0:
1. Place the cuff on the arm (preferably the left arm).
2. Press and hold the
number about 30 to 40 mmHg higher than
your expected systolic pressure appears.
3. When the desired number is reached, release
the
button to start measurement.
START
Continue to measure your blood pressure as
described on the previous page.
When the memory setting is 100:
1. Place the cuff on the arm (preferably the left arm).
2. Press the
START
All of the display segments are displayed.
Zero (0) is displayed blinking briefly.
3. Press and hold the
until a number about 30 to 40 mmHg
higher than your expected systolic
pressure appears.
4. When the desired number is reached,
release the
measurement. Continue to measure
your blood pressure as described on
the previous page.
Measurements
button until a
START
button.
START
button to start
START
Release the button
button
English 13
At heart level
Press and hold
the button
at the desired
systolic pressure
Refer to the previous
page for measurement
At heart level
Press
All of the display
segments displayed
Zero blinking
Press and hold
the button
Release the button
at the desired
systolic pressure
Refer to the previous page
for measurement
START
START
START