Bayer HealthCare Contour next USB User Manual page 41

Blood glucose monitoring system
Hide thumbs Also See for Contour next USB:
Table of Contents

Advertisement

Accuracy
The C
®
N
USB blood glucose monitoring
ONTOUR
EXT
system was tested on 100 capillary blood samples using
C
®
N
test strips. Two replicates were tested with
ONTOUR
EXT
each of three lots of C
600 readings. Results were compared to the YSI
analyzer, which is traceable to the CDC hexokinase method.
The tables below compare the performance of the two
methods.
Table 1 — System accuracy results for C
meter using C
ONTOUR
concentration < 4.2 mmol/L.
Difference range in values
between YSI laboratory
reference method and
C
N
USB meter
ONTOUR
EXT
Number (and percent)
of samples within
specified range
Table 2 — System accuracy results for C
meter using C
ONTOUR
concentration ≥ 4.2 mmol/L.
Difference range in values
between YSI laboratory
reference method and
C
N
USB
ONTOUR
EXT
meter
Number (and percent)
of samples within
specified range
Acceptance criteria in ISO 15197:2003 are that 95% of all
differences in glucose values (i.e., between reference method and
meter) should be within 0.8 mmol/L for glucose values less than
4.2 mmol/L, and within 20% for glucose values greater than or equal
to 4.2 mmol/L.
74
84318354_CntrNext_USB_UG_ENca_FpBp_v1.indd 74-75
84318354_CntrNext_USB_UG_ENca_FpBp_v1.indd 74-75
N
test strips for a total of
ONTOUR
EXT
®
glucose
N
USB
ONTOUR
EXT
N
test strips at glucose
EXT
Within ±
Within ±
Within ±
0.28
0.56
0.83
mmol/L
mmol/L
mmol/L
65 / 78
78 / 78
78 / 78
(83.3%)
(100%)
(100%)
N
USB
ONTOUR
EXT
N
test strips at glucose
EXT
Within ±
Within ±
Within ±
Within ±
5%
10%
15%
20%
390 / 522
512 / 522
522 / 522
522 / 522
(74.7%)
(98.1%)
(100%)
(100%)
Precision
A repeatability study was conducted with the
C
®
N
USB blood glucose monitoring system using
ONTOUR
EXT
100 venous whole blood specimens with glucose levels from
2.6 to 18.4. Multiple replicates (n=300) were tested using
multiple C
N
USB blood glucose meters and
ONTOUR
EXT
®
three lots of C
N
blood glucose test strips. The
ONTOUR
EXT
following precision results were obtained.
System repeatability results for C
meter using C
N
test strips
ONTOUR
EXT
Pooled Standard
Mean, mmol/L
Deviation
2.6
0.05
4.8
0.07
7.3
0.11
11.7
0.15
19.0
0.35
N
USB
ONTOUR
EXT
Coefficient of
Variation, %
1.8
1.5
1.5
1.3
1.9
75
6/26/12 10:29 AM
6/26/12 10:29 AM