SECTION 3—CHECKLIST AND CYLINDER INSPECTION
SECTION 3—CHECKLIST AND
CYLINDER INSPECTION

Compressor Operation Checklist

Each time the HomeFill II Compressor is used to fill a
cylinder, complete the following checklist:
Ensure the concentrator has been On for at least thirty
minutes. Refer to the concentrator Owner's Manual.
Perform the prefill inspection on the cylinder.
Connect the cylinder to the compressor.
Push the compressor power switch to the on ( ) position.
Examine the indicator lights on the control panel.
Disconnect and remove the full cylinder.
Push the compressor power switch to the Off (O) position.
If filling another cylinder, repeat this checklist.

Cylinder Prefill Inspection

ONLY use cylinders that have the Invacare HomeFill II
connection and the label.
All cylinders MUST be inspected before attempting to fill.
Otherwise, injury or damage may occur.
NOTE: For this procedure, refer to FIGURE 3.1 on page 17.